No more He in Mono Hutch - What to do ?
In the Monochromator hutch:
We use a liquid He vessel to have He flux in Mono chamber | He Level in Mono Hutch - Visualization on Monitor in Control Room |
Ball must be between 2 marks
Ball down means there is NO MORE He in Mono chamber. Stop immediately the beam! Danger for Be window !!!!
To change He bottle:
1) First thing to do: Close mechanical valve 0 in Experimental hutch |
2) Close mechanical valve 1 in Mono hutch | |
|
||
3) Close mechanical valve 3 in Mono hutch | 4) You are ready to pump Mono chamber | |
Connect pumping group to Mono chamber |
||
5) Verify that Mono Hutch Electronic Valves are open | 6) Open completely mechanical valve 2 in Mono hutch to pump the whole chamber | |
Image 'vacuum synopsis' |
|
|
7) Call Floor coordinator (25-25) to get a new liquid He bottle Connect it - Note new date on camera field of view 9) Close mechanical valve 2 - Stop pumping
11) Do not forget to reopen mechanical valve 0 in Experimental Hutch before working
|
8) Check vacuum state on jauge 10) Open mechanical valve 1, then 3 He surpression must go up the ball |