Comparison IXS versus INS
Similarities and differences between the IXS and INS cross section are summarised below:
-
Q - E limitation for INS
-
Small beam size for IXS
- X-rays couple to the electrons of the system with a cross-section proportional to the square of the classical electron radius, ro=2.82.10-13cm, i.e. with a strength comparable to the neutron-nucleus scattering cross-section b.
- The IXS cross section is proportional to fj(Q)2. In the limit Q→0, the form factor is equal to the number of electrons in the scattering atom, Z; for increasing values of Q, the form factor decays with decay constants of the order of the inverse of the atomic wavefunction dimensions of the electrons in the atom.
- The total absorption cross-section of X-rays above 10 keV energy is limited in almost all cases (Z>4) by the photoelectric absorption process, and not by the Thomson scattering process. The photoelectric absorption, whose cross-section is roughly proportional to Z4, determines therefore the actual sample size along the scattering path. Consequently the Thomson scattering channel is not very efficient for system with high Z in spite of the Z2 dependence of its cross-section.
- As a consequence of the above point multiple scattering processes can in general be neglected
- The magnetic cross section is negligible for IXS, whereas it is comparable to the nuclear cross section for neutrons.
- The IXS cross section is highly coherent
- The shape of the IXS instrumental energy resolution is not Gaussian as it is for a neutron triple-axis spectrometer, but Lorentzian.