In some cases, the Dozor plot and/or characterization results indicate improper centering of the crystal. 

Screenshot 2024-10-23 at 8.02.32 PM.png

This can come from four situations:

  1. Crystal not centered on the rotation axis
  2. Anisotropic diffraction of the crystal
  3. Beam position not at the center of the screen
  4. Rotation axis miscalibrated

If you are sure that situations one and two do not apply (usually situation 1 can be ruled out y using Xray centering or MXPress workflows), you can align the beam and calibrate the rotation axis. NB: if you have a small feature that you are trying to 3-Click center on but cannot, the problem is #4.

The process for centering the beam and calibrating the axis is very simple but requires a diffracting sample:

  1. mount a diffracting sample (preferably one that has already been analyzed)
  2. Run the "calibrate_rotation_axis" workflow

NOTE: Two quality control checks are performed in this procedure which prevent the rotation axis position from being updated if

  1. The position has moved very far (currently 25 microns, but being refined) 
  2. The correlation between the 0 and 180 degree scans is less than 80%

Both issues could be due to a poorly diffracting sample, so it would be worth selecting one from the test puck (contact your local contact) or a sample you know to diffract well (2A, single lattice).  What this process actually does:

  1. Center beam (the same as in Beamline actions->"Centre Beam" (which can ad should be done independetly of the the calibrate_rotation_axis workflow
  2. Does two large line scans 180 degrees apart
  3. Determines the shift (if any between them)
  4. Sets this alignment motor position to be the reference position in Mxcube and the Microdiff software

A video can be found here, which includes the now unnecessary step of doing a center beam beforehand (center beam has been included in the workflow as of Nov 2024)

Beam alignment, then diffractive gonio alignment