Vacuum not good
Vacuum troubles
All beamline components except detector are in the vacuum: very good one to protect Optical elements until the Beryllium window and a bit less good until the Kapton window at the end of the flight tube. There are few tips what can be wrong if some vacuum reading makes experiment impossible.
No more vacuum upstream of Experimental hutch (after Be-window) and big primary pump in fault state
The big primary pump at the end of the Optics hutch needs for its operation cooled water. In the case of a water cut, the pump should be preferably shut down. If not done it will stop in a fault state. To make it working back (once water is circulating again), one should first stop it which will erase fault state and that put ON again. And wait...until vacuum reach 10-4 level to continue the experiment.